Infusion pump controller
Volumetric pump drive with occlusion and air-in-line detection.
Function
Drives a peristaltic or cassette pump at a programmed flow rate, watches downstream pressure and air-in-line sensors, and stops and alarms on occlusion, free flow or a door opening.
Design choices
- Two independent processors: one controls the motor, the other supervises and can cut motor power.
- Closed-loop motor control with encoder feedback to hold flow accuracy across the rate range.
- Ultrasonic air-in-line sensor and a strain-gauge pressure channel with self-test on every start.
- Battery backup with a fuel gauge so infusions continue through mains loss.
Schematic
Image coming soon
PCB
Image coming soon
Future challenges
- Proving the safety architecture to regulators (IEC 60601, IEC 62304).
- Reducing nuisance alarms without missing real occlusions.
- Securing drug-library updates delivered over hospital networks.